2.6 Parameters, key <PARAM>

751 GPD Titrino

35

stop crit: drift
**titr.
stop drift 20 ul/min
**titr.

t(delay) 10 s
**titr.

stop time OFF s
**titr.

Type of stop criteria (drift, time)

Titration stops if EP and stop drift are reached (1...999

uL/min)

Titration stops if there is no dosing during t(delay)

(0...999 s, INF)

<CLEAR> sets "INF"

Switch off when EP is reached and the set time after

the last dosing has elapsed.

If "INF" = an inquiry for the stop time appears.

Stop after a time (0...999 999 s, OFF)

<CLEAR> sets "OFF".

Stop after the set time after start of titration. "OFF"

means no stop, i.e. titration for an "infinitely" long time.

>titration parameters

titr.direction: -

pause 1 0 s

**titr.

start V: OFF
cond.

start V 0.0 ml
cond.

factor 0
cond.

dos.rate max. ml/min
**titr.

Titration parameters

Direction is set automatically (+, -, auto)

auto: The direction is set automatically by the Titrino.

+:

Direction of higher voltage (more "positive"),

larger currents.

- :

Direction of lower voltage, smaller currents.

Waiting time before start volume (0..999 999 s)

Waiting time, e.g. for equilibration of the electrode after

start. The waiting time can be aborted with <QUIT>.

Type of start volume (OFF, abs., rel.)

"OFF": start volume switched off

"abs.": absolute start volume in mL

"rel.":

relative start volume to sample size.

If "abs." is set:

Absolute start volume (0...999.99 mL)

If "rel." is set:

Factor for relative start volume (0...±999 999).

Calculated as: start V in mL = factor * sample size

Dosing rate for start volume (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min