2.6. Parameters, key <PARAM>

751 GPD Titrino

48

rate: OFF
**titr.

low lim. 0.000 ml/min
up lim. 150 ml/min
action: none
**titr.

temperature: OFF
**titr.

low lim. -170.0 °C
up lim. 500.0 °C
action: none
**titr.

assign output L10: none

violated limit: any

output L10: pulse

Monitoring of rate (ON, OFF)

With "on" follow the requests:

Limit values (0.000...150 mL/min)

Action if a limit is exceeded (end, hold, wait, none)

end:

Abort.

hold:

Hold reagent addition until manually restarted.

wait:

Hold reagent addition until limits are again

complied with, then automatically continue. If

the lower limit has been exceeded, "wait" is

senseless as the rate will only become smaller

during the waiting time-

Monitoring of temperature (ON, OFF)

With "on", the following requests:

Limit values (-170.0...500.0°C)

Action if a limit is exceeded (end, hold, wait, none)

end:

Abort.

hold:

Hold reagent addition until manually restarted.

wait:

Hold reagent addition until limits are again

complied with, then continue automatically.

Assignment of output L10 for monitoring (meas, temp,

rate, all, none)

A signal can be output at L10 (pin 8) of the remote

socket if a limit has been violated.

Assignment of a violated limit (any, upper, lower)

The signal will be set if the assigned limit is violated.

Type of signal to be set when limits are violated. (active,

pulse)

active: Output line is set to 0 V.

pulse: Pulse > 100 ms.

Important: An active line will be set inactive with an

impulse!

Same possibilities for lines L11, L12, L13.

Pin assignment on socket "Remote":

L10

Pin 8

L11

Pin 13

L12

Pin 19

L13

Pin 20