2.6 Parameters, key <PARAM>

751 GPD Titrino

69

2.6.7 Parameters for MEAS

measuring parameters
determine the measurement.
statistics:
Calculation of mean values and standard deviation,
see page 77.
preselections:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.

>measuring parameters

signal drift OFF mV/min

equilibr.time OFF s

meas.input: 1

I(pol) 1 uA

U(pol) 400 mV

electrode test: OFF

Measuring parameters

Drift criterion for measured value acquisition (input

range depends on the measured quantity:

pH, U, Ipol:

0.5...999 mV/min, OFF

Upol:

0.05...99.9 uA/min, OFF

T:

0.5...999 °C/min, OFF)

<CLEAR> sets "OFF".

"OFF" means that the measured value is acquired after

a fixed equilibration time.

Waiting time for measured value acquisition (0...9999 s,

OFF)

<CLEAR> sets "OFF".

If no new equilibration time has been entered, the

Titrino calculates an equilibration time appropriate to

the drift with the formula

150

equilibr.time (in s) = ДДДДДДДДД + 5

ûDrift + 0.01

The measured value is acquired when the first criterion

(drift or time) is met. With drift and time "OFF", you will

have an "infinite" measurement.

Measuring input for pH and U. (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 215.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 uA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby mode to a

measurement. "OFF" means that the test is not

performed.

parameters

>measuring parameters

>statistics
>preselections