Metrohm 751 GPD Titrino User Manual

Page 117

Advertising
background image

3.1. General rules

751 GPD Titrino

114

E171

Rate too low in DOS. No dispensing possible with the Exchange Unit currently

mounted.

Exit: The error message disappears on next start or &m $S.

E172

In TIP or DOS a QuickMeas was started, without defining a measuring quantity.

Exit: The error message disappears on next start or &Mode.QuickMeas $S.

E173

The warning interval of the internal buret D0 called.

Exit: Execute prep &a.b.p $G or start next method.

E174

The warning interval of the external buret D1 called.

Exit: Execute prep &a.b.p $G or start next method.

E175

The warning interval of the external buret D2 called.

Exit: Execute prep &a.b.p $G or start next method.

E176

The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was interrupted

manually.

Exit. The error message disappears on next start.

E177

Accessing to the memory card, the card was not (properly) inserted.

Exit. The error message disappears on next start.

E178

The date of changing the battery of the card is expired.

Exit. The error message disappears on next start.

E180

Memory card write-protected.

Exit: Send new command.

E181

Memory card not formatted.

Exit: Send new command.

E182

Memory card not accessible.

Exit: Send new command.

E183

A directory with the same name exists already on the memory card.

Exit: Send new command.

E184

Measured value below lower limit.

Exit: The error message disappears when again within the limits or &m $S.

E185

Measured value above upper limit.

Exit: The error message disappears when again within the limits or &m $S.

E186

Temperature below lower limit.

Exit: The error message disappears when again within the limits or &m $S.

E187

Temperature above upper limit.

Exit: The error message disappears when again within the limits or &m $S.

E188

Rate below lower limit.

Exit: The error message disappears when again within the limits or &m $S.

E189

Rate above upper limit.

Exit: The error message disappears when again within the limits or &m $S.

E196

Result is out of limits.

Exit: The error message disappears on next start or on recalculation.

E197

Sample size is out of limits.

Exit: The error message disappears on next start or on introduction of new

sample size.

E198

Validation interval is expired.

Exit: The error message disappears on next start or clear counter with

&Config.Monitoring.Validation.ClearCount $G.

E199

Service date is reached.

Exit: The error message disappears on next start or change date in

&Config.Monitoring.Service.Date.