2.6 Parameters, key <PARAM>

751 GPD Titrino

51

Control parameters

Dosing outside the control range:

The dosing rate is given mainly by the

parameter "max.rate".

Optimize the parameters "max.rate"

and "dynamics" together to avoid

overtitration the first time the control

point is reached.

Set "dynamics" such that the

measured value lies inside the control

range during the phase where the

value is kept. Set "dynamics" to e.g.

pH=3, U=180 mV for slow reactions.

You can also make a SET pretitration

to reach the control point. In this case,

"dynamics" of the STAT titration is not

very crucial.

Dosing inside the control range:

Here the dosing rate is given mainly by

the parameter "min.rate".

Rule of thumb for "min.rate" in ul/min

= (expected rate of the reaction in

ul/min)/10.

How to proceed if ...

Problem

Possible causes and corrective action

Adherence to the control point is

not good. The measured value

is too high, then too low. "The

controller oscillates".

Set "max.rate" and possibly also "min.rate" lower.

Set "dynamics" larger.

Is the stirring efficiency adequate?

Arrange electrode and buret tip properly, see page

217.

Are the tubing lengths kinked?

Possibly use Exchange Unit with a small cylinder

volume (smaller volume increment per pulse).

Time for reaching the control

point is too long.

Set "dynamics" smaller.

Set "min.rate" higher.

Set "max.rate" higher.

U/mV

V/ml

Control

range

EP

t/s

V/ml