3.2. Remote control commands

751 GPD Titrino

166

With "ON", executed lines are copied to the next free silo lines, see page 96. Ex-
ercise caution if you edit the silo memory during the determinations!

3.2.2.

3.2.2.128.

128. SSmplData.ON

.ONSilo.S

.SaveLines

ON, OFF

OFF

Silo lines are not deleted when they are worked off. Assigned results are stored
as C24 and C25. "Save lines" can only be set to "ON" if the silo is completely
empty. Delete the silo, see 3.2.2.126.

3.2.2.

3.2.2.129.

129. HHotKey.U

.User.N

.Name

up to 10 ASCII characters

HHotKey.U

.User.D

.Delete

$G

HHotKey.U

.User.D

.Delete.N

.Name

up to 10 ASCII characters

HHotKey.U

.User.DelA

.DelAll

$G

HHotKey.U

.User.L

.List.1.N

.1.Name

read only

Management of user names.
.Name:

Input of user names.

.Delete.Name: Deletes selected user name with &HotKey.User.Delete $G.
.List:

List of all user names.

3.2.2.

3.2.2.130.

130. IInfo.R

.Report

$G

IInfo.R

.Report.S

.Select

configuration, parameters, smpl data,

statistics, silo, calib, C-fmla, def, user method, full

full, short, mplist,

curve, deriv, comb, meas crv, temp crv, adj para, scalc full,

scalc srt, prep, calc, act dir, mem card, all, ff

$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running determi-

nation.

parameters: Parameter report of the current method. During a running determi-

nation only "live"-parameters are accessible.

smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo:

Contents of the silo memory.

calib:

Calibration data of the measuring input in the current method.

C-fmla:

Contents of the <C-fmla> key.

def:

Contents of the <def> key.

user method: Contents of the method memory.
full:

Full result report of the last completed determination.

short:

Short result report of the last completed determination.

mplist:

Measuring point list of the running determination.

curve:

Titration curve of the last determination.

derive:

1st derive of titration curve of the last determination (with DET).

comb:

1st derive combined with the titration curve of the last determina-
tion (with DET).

meas crv: Curve measured value vs. time (with STAT, DOS, DOC and acti-

vated meas monitoring).

temp crv: Curve temperature vs. time (with STAT, DOS, DOC and activated

temperature monitoring).

adj para: Adjustment parameters.
scalc full: Full report of the silo calculations.