Metrohm 751 GPD Titrino User Manual

Page 203

Advertising
background image

4.2. Diagnosis

751 GPD Titrino

200

Press <5>.

RS232 test 1...2

8.1 RS232 test 1

Press <1>.

RS232 test-connector? 1

Insert the 3.496.8560 test plug in port ‘A1’.

Press <Enter>.

Test sequence:

1. In case of an error an error message is displayed. If for example no test plug is

connected, error message

error 68 appears).

2. If no error is found, after about 5 s display shows

RS232 test 1 o.k.

Remove test plug.

Press <Clear>.

RS232 test 1...2

8.2 RS232 test 2

Press <2>.

RS232 test-connector? 2

Insert the 3.496.8560 test plug in port ‘A2’.

Press <Enter>.

Test sequence:

1. In case of an error an error message is displayed. If for example no test plug is

connected, error message

error 68 appears)

2. If no error is found, after about 5 s display shows

RS232 test 2 o.k.

Remove test plug.

Press <Clear>.

RS232 test 1...2

Press <Clear>.

diagnose press key 0...9