3.2. Remote control commands

751 GPD Titrino

124

*Parameter Tree part "Parameters for DOS"

Ã

.

DosPara

Dosing parameters

³ Ã

.Type

Type of dosing

volume, time, rate

3.2.2.52.

³ Ã

.Volume

Volume dosing

³ ³ Ã

.Volume

Volume

0...10...99 999.99

ditto

³ ³ Ã

.DisType

Second dosing criterion

time, rate

ditto

³ ³ Ã

.Rate

Rate

0.001...150, max.

ditto

³ ³ Ã

.Time

Dosing time

1...100...999 999

ditto

³ Ã

.Time

Time dosing

³ ³ Ã

.Time

Dosing time

1...100...999 999

ditto

³ ³ Ã

.DisType

Second dosing criterion

volume, rate

ditto

³ ³ Ã

.Rate

Rate

0.001...150, max.

ditto

³ ³ Ã

.Volume

Volume

0...10...99 999.99

ditto

³ Ã

.Rate

Rate dosing

³ ³ Ã

.Rate

Rate

0.001...150, max.

ditto

³ ³ Ã

.StopType

Second dosing criterion

volume, time

ditto

³ ³ Ã

.Time

Dosing time

1...100...999 999

ditto

³ ³ Ã

.Volume

Volume

0...10...99 999.99

ditto

³ Ã

.Pause

Waiting time after start volume

0...999 999

3.2.2.53.

³ Ã

.TDelta

Time interv.for meas.acquisition

1...10...999 999

3.2.2.54.

³ Ã

.DosUnit

Selection of the dosing unit

internal D0, external D1,

³ ³

external D2

3.2.2.17.

³ Ã

.Temp

Titration temperature

-170.0...25.0...500.0

3.2.2.55.

³ ³

Ã

.

StopCond

Stop conditions

³ Ã

.VStopStop volume

³ ³ Ã

.Type

Type of stop volume

abs., rel., OFF

3.2.2.20.

³ ³ Ã

.V

Volume for absolute stop volume

0...99.99...9999.99

ditto

³ ³ Ã

.Factor

Factor for relative stop volume

0...

±

999 999

ditto

³ Ã

.FillRate

Filling rate

0.01...150.0, max.

3.2.2.23.

³ ³

Ã

.

StatisticsStatistics

³ Ã

.Status

Status of statistics calculation

ON, OFF

3.2.2.24.

³ Ã

.MeanN

No.of individual determinations

2...20

ditto

³ Ã

.ResTab

Result table

³ ³ Ã

.Select

original,delete n,delete all

ditto

³ ³ Ã

.DelN

Deletion of individual results

1...20

ditto

³ ³

Ã

.

Monitoring

Monitoring

³ Ã

.MeasVal

Monitoring of measured values

³ ³ Ã

.MeasMode

Measured value

pH, U, Ipol, Upol, OFF

3.2.2.56.

³ ³ Ã

.MeasInput

Measuring input

1, 2, diff.

ditto

³ ³ Ã

.PCurrent

Polarization current

0...1...

±

127

ditto

³ ³ Ã

.PVoltage

Polarization voltage

0...400...

±

1270

ditto

³ ³ Ã

.PolElectrTest

Test for polarized electrodes

ON, OFF

ditto

³ ³ Ã

.pH

Limits for monitoring of pH values

³ ³ ³ Ã

.LowLim

Lower limit

-20.00...20.00

ditto

³ ³ ³ Ã

.UpLim

Upper limit

-20.00...20.00

ditto

³ ³ Ã

.U

Limits for monitoring of U values

³ ³ ³ Ã

.LowLim

Lower limit

-2000...2000

ditto

³ ³ ³ Ã

.UpLim

Upper limit

-2000...2000

ditto

³ ³ Ã

.Ipol

Limits for monitoring of Ipol values

³ ³ ³

:

as for U

³ ³ Ã

.U

Limits for monitoring of Upol values

³ ³ ³

:

as above

³ ³ Ã

.Action

Action if out of limits

end, hold, wait, none

ditto