Metrohm 751 GPD Titrino User Manual

Page 182

Advertising
background image

3.2 Remote control commands

751 GPD Titrino

179

.Right:

Relative specification of the width of the output medium (e.g. paper
width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.

.Feed:

Length of the volume axis referred to the burette cylinder volume,
V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange
Unit). Depending on the printer, the measure in cm may not always
be correct.

3.2.2.

3.2.2.171.

171. Se

Setup.P

.PowerOn

$G

Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.

3.2.2.

3.2.2.172.

172. Se

Setup.I.Initialise

$G

Se

Setup.I.Initialise.S

.Select

ActMeth

ActMeth, Silo, Calib, Config,

Assembly, Setup, All

Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for the

data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Calib:

pH calibration data for all measuring inputs.

Config:

All values under &Config.

Assembly: All values under &Assembly.
Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.

3.2.2.

3.2.2.173.

173. Se

Setup.R

.RamInit

$G

Initializes instrument, see page 204. All parameters are set to their default value
and error messages are cleared. The user and silo memories will be deleted.
The user memory contains the default user methods from Metrohm.

3.2.2.

3.2.2.174.

174. Se

Setup.Ins

.InstrNo

$G

Se

Setup.Ins

.InstrNo.V

.Value

serial number

serial number, 8 ASCII characters

Instrument identification for report output.
Set the value with &Setup.InstrNo $G .

3.2.2.

3.2.2.175.

175. DDiagnose.R

.Report

$G

Output of the report containing the adjustment parameters. The Titrino has to be
in its inactive basic state.