3.2. Remote control commands

751 GPD Titrino

148

.ResTab.DelN: Specification of the line number to be deleted.

3.2.2.

3.2.2.25.

25. M

Mode.P

.Parameter.E

.Evaluation.EEPC

DET: 0...55...200
MET pH: 0.1...0.50

0.50...9.99

U, Ipol: 1...30

30...999

Upol:

0.1...22...99.9

M

Mode.P

.Parameter.E

.Evaluation.R

.Recognition.S

.Selected

all

all, greatest,

last, window, OFF

M

Mode.P

.Parameter.E

.Evaluation.R

.Recognition.W

.Window.1.L

.1.LowLim

pH:

0...

±

20.00, OFF

OFF

U, Ipol:

0...

±

2000, OFF

OFF

Upol:

0...

±

200.0, OFF

OFF

M

Mode.P

.Parameter.E

.Evaluation.R

.Recognition.W

.Window.1.U

.1.UpLim

Input range as LowLim

etc. up to 9 windows
M

Mode.P

.Parameter.E

.Evaluation.F

.FixEP.1.V

.1.Value pH:

0...

±

20.00, OFF

OFF

U; Ipol: 0...

±

2000, OFF

OFF

etc. up to 9 fix EP's

Upol: 0...

±

200.0, OFF

OFF

M

Mode.P

.Parameter.E

.Evaluation.p

.pK

ON, OFF

OFF

Parameters for DET and MET: Evaluation of the EP's, see page 21.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).
.Recognition.Selected: EP recognition.

all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.

.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U and

Ipol) or uA (with Upol).

.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and

Ipol) or uA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 23.

.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. uA (for Upol). Fix EP's are

evaluated until the setting OFF is found.

.pK: pK or HNP evaluation. Possible only in pH and U titrations.

3.2.2.

3.2.2.26.

26. M

Mode.P

.Parameter.P

.Presel.I.IReq

id1, id1&2, all, OFF

OFF

M

Mode.P

.Parameter.P

.Presel.S

.SReq

value, unit, all, OFF

OFF

Parameters for DET, MET, SET, KFT, STAT, DOS, DOC, MEAS: Automatic in-
quiry after the start of the determination. From such an inquiry, the determina-
tion continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.122) or with &M $G, see 3.2.2.1.
$H is not possible in requests.

3.2.2.

3.2.2.27.

27. M

Mode.P

.Parameter.P

.Presel.L

.LimSmplSize.S

.Status

ON, OFF

OFF

M

Mode.P

.Parameter.P

.Presel.L

.LimSmplSize.L

.LoLim

0.0

0.0...999 999

M

Mode.P

.Parameter.P

.Presel.L

.LimSmplSize.U

.UpLim

0.0...999 999

999 999

Limit control for the sample size.