2.10 Data output

751 GPD Titrino

81

scalc full

Full report of silo calculations.

scalc srt

Short report of silo calculations.

param

Parameter report.

calc

Report with formulas and operands.

calib

Calibration data.

ff

Form feed on printer.

Original reports which are put out automatically at the end of the titration can be printed

with recalculated values at any time. Key sequence:

<PRINT><REPORTS><ENTER>.

Original reports have double dashes ==== at the end, whereas recalculations are

marked by single dashes ----.

Report outputs can be stopped with <QUIT>.

Example of reports:

Full result report

Card label if a card was inserted, see page

92.

Device label (if there is a designation) and

manual signature

Titration curve

Scaling of volume and pH axis

'fr
751 GPD Titrino 0P1/101 751.0020
date 1999-06-27 time 08:54 14
card label: Appl.751
pH(init) 2.88 DET pH ********
smpl size 0.372 g
EP1 2.083 ml 7.64
RS1 3.80 g/l
device label Titr 1 sign:
============

'cu
751 GPD Titrino 0P1/101 751.0020
date 1999-06-27 time 08:54 14
start V 0.000 ml DET pH ********
1.0 ml/div dpH=1.0/div

============