Wrong sample. For SET, KFT or DOC with preset titration direction

the first measured value lies behind the endpoint.

Exit: The error message disappears on next startup.

E131

No EP set for SET, STAT.

Exit: The error message disappears on next startup.

E132

Silo empty and it has been started with open silo or empty silo has

been opened.

Exit: Send a silo entry.

E133

Silo full.

Exit: Send new command.

E134

No method. A method, which is required from the silo memory or in TIP, does

not exist.

Exit: The error message disappears on next startup.

E135

Check temp.sensor in MEAS T or with activated temperature monitoring.

Exit: Correct error or &m $S.

E136

Same buffer in CAL. Measured value of the second buffer differs less than 6

mV from the measured value of the first buffer.

Exit: Correct error or &m $S.

E137

XXX Bytes are missing so that the method, the silo line could not be stored or

not enough RAM for running TIP.

Exit: Send new command.

E149

Rate missing for calculation.

Exit: The error message disappears on next start or on recalculation.

E150

Number of rates does not agree with the set windows or the mean rate C80

could not be calculated.

Exit: The error message disappears on next start or on recalculation.

E151

Fix volume missing for calculation.

Exit: The error message disappears on next startup or on recalculation.

E155

No new silo result (C24 or C25).

Exit: The error message disappears on next start or on recalculation.

E157

No sequence defined in TIP.

Exit. The error message disappears on next start.

E158

A second TIP has been called up in TIP.

Exit: The error message disappears on next start.

E160

No new temporary variable.

Exit: The error message disappears on next start.

E161

Measurement range of the secondary measured value (temperature) exceeded.

The primary measured value (pH, U, Ipol, Upol) can also be unstable.

Exit: Rectify error or &m $S.

E162

No ramp end set in DOC.

Exit: The error message disappears on next start or &m $S.

E166

Save lines is "OFF" although a submethod of TIP includes an assignment to

C24 or C25.

Exit: The error message disappears on next start. Attention: The data of this

sample will not be stored.

E167

Rate too high in DOS. No dispensing possible with the Exchange Unit currently

mounted.

Exit: The error message disappears on next start or &m $S.

E170

Fix time missing for calculation.

Exit: The error message disappears on next start or on recalculation.