Metrohm 751 GPD Titrino User Manual

Page 35

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

32

Relation between the stop criteria "time" and "drift"

The stop criterion "time", t(delay), means that the end point must be exceeded for a

certain period of time. In other words, after the last dosed increment, time t is allowed to

elapse before the titration is stopped. The size of this last increment depends on the

volume of the Exchange Unit used. With a 20 mL Exchange Unit, the smallest possible

increment is 2 ul. With a t(delay) = 5 s, the last 2 ul reagent must thus suffice for 5 s or

longer. This results in a drift of

2 uL/5 s = 24 uL/min (the drift can be less than 24

uL/min as it is not known whether the last increment would also suffice for 10 s). If you

have been working up to this point with a 20 mL Exchange Unit and a t(delay) = 5 s, you

can set a value of

24 uL/min as stop drift. The following table shows several values for

the maximum stop drift.

t(delay)

min.incr. (Exchange Unit)

5 s

10 s

20 s

0.5 uL (5 ml)

6 uL/min

3 uL/min

1.5 uL/min

1 uL (10 ml)

12 uL/min

6 uL/min

3 uL/min

2 uL (20 ml)

24 uL/min

12 uL/min

6 uL/min

5 uL (50 ml)

60 uL/min

30 uL/min

15 uL/min

Same t(delay) with a range of extremely small volume increments means different switch-

off points. In case the stop criterion "drift" is used, the stopping point remains the same.

Advertising