Metrohm 751 GPD Titrino User Manual

Page 160

Advertising
background image

3.2 Remote control commands

751 GPD Titrino

157

3.2.2.

3.2.2.70.

70. M

Mode.P

.Parameter.C

.Calibration.S

.SignalDrift

0.5...22...999, OFF

M

Mode.P

.Parameter.C

.Calibration.E

.EquTime

0...110

110...9999, OFF

Parameters for CAL: Criteria for measured value acquisition. Measured value
drift in mV/min, equilibration time in s. OFF means that the corresponding crite-
rion is switched off. If both criterions are on OFF, the measured value is ac-
quired immediately.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 20. After it has been edited once, it
remains in force with the set value.

3.2.2.

3.2.2.71.

71. M

Mode.P

.Parameter.C

.Calibration.El

.ElectrodeId

up to 8 ASCII char

Parameters for CAL: Electrode identification. It is classified under calibration
data, see 3.2.2.131.

3.2.2.

3.2.2.72.

72. M

Mode.P

.Parameter.C

.Calibration.Sm

.SmplChanger

ON, OFF

OFF

Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.

3.2.2.

3.2.2.73.

73. M

Mode.P

.Parameter.C

.Calibration.A

.ActPulse

first, all, OFF

OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page 224.

3.2.2.

3.2.2.74.

74. M

Mode.P

.Parameter.S

.Sequence.X.S

.X.Select

method, pause, L4 output,

L6 output, info, prep, stirrer, OFF

OFF

Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.75.

3.2.2.

3.2.2.75.

75. M

Mode.P

.Parameter.S

.Sequence.X.M

.X.Method

Method name

M

Mode.P

.Parameter.S

.Sequence.X.P

.X.Pause

00...999 999, INF

M

Mode.P

.Parameter.S

.Sequence.X.L

.X.L4Output

active, inactive,

pulse, OFF

OFF

M

Mode.P

.Parameter.S

.Sequence.X.L6

.X.L6Output

as for L4

M

Mode.P

.Parameter.S

.Sequence.X.I

.X.Info

up to 16 ASCII characters

M

Mode.P

.Parameter.S

.Sequence.X.Pr

.X.Prep

internal D0

internal D0, external D1,

external D2

M

Mode.P

.Parameter.S

.Sequence.X.S

.X.Stirrer

ON

ON, OFF

Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the

card. Up to 8 ASCII characters.

.Pause:

Pause time in s. INF means infinite. Continue the sequence with &m
$G.

.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin

3) in a submethod sets an output set to active in TIP to inactive.