3.2 Remote control commands

751 GPD Titrino

161

Monitoring of validation.
.Interval:

Time interval in days for validation.

.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.

3.2.2.

3.2.2.92.

92. CConfig.M

.Monitoring.C

.Calibration.S

.Status

ON, OFF

OFF

CConfig.M

.Monitoring.C

.Calibration.M

.MeasInput

11, 2, diff

CConfig.M

.Monitoring.C

.Calibration.I.Interval

1...77...9999

CConfig.M

.Monitoring.C

.Calibration.C

.Counter

00...9999

Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval:

Time interval in days for calibration from the last calibration date.

.Counter: Time counter in days since last calibration. The counter is reset to

zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.

3.2.2.

3.2.2.93.

93. CConfig.M

.Monitoring.S

.Service.S

.Status

ON, OFF

OFF

CConfig.M

.Monitoring.S

.Service.D

.Date

XXXX-XX-XX

Monitoring of service interval.

3.2.2.

3.2.2.94.

94. CConfig.M

.Monitoring.D

.DiagRep

ON, OFF

OFF

Printing of system test report after each switching on of the Titrino.

3.2.2.

3.2.2.95.

95. CConfig.P

.PeriphUnit.C

.CharSet1

Epson, Seiko, Citizen, HP, IBM

IBM

CConfig.P

.PeriphUnit.CharSet2

.CharSet2

Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.

3.2.2.

3.2.2.96.

96. CConfig.P

.PeriphUnit.R

.RepToComport

11, 2, 1&2

Selection of COM of the Titrino where manually triggered reports should be out-
putted.

3.2.2.

3.2.2.97.

97. CConfig.P

.PeriphUnit.B

.Balance

Sartorius

Sartorius,Mettler,Mettler AT,

AND,Precisa

Selection of the balance type.

3.2.2.

3.2.2.98.

98. CConfig.P

.PeriphUnit.S

.Stirrer

ON, OFF

OFF

Automatic stirrer control. With "ON" the stirrer will be switched on after starting
a method. At the end of the method it is switched off again. Valid for DET, MET,
SET, KFT, STAT, DOC.

3.2.2.

3.2.2.99.

99. CConfig.P

.PeriphUnit.Rem

.RemoteBox.S

.Status

ON, OFF

OFF