Metrohm 751 GPD Titrino User Manual

Page 67

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

64

dos.element: internal D0

meas.input: 1

I(pol) 1 uA

U(pol) 400 mV

electrode test: OFF

temperature 25.0 °C

Selection of dosing element (internal D0, external D1,

external D2)

internal D0: internal buret of the Titrino

external D1/2: buret D1, D2, resp.

Measuring input (1, 2, diff.)

Measuring input for pH and U. Measuring input 1 or 2

or differential amplifier; connection of electrodes, see

page 215.

With polarized electrodes, the request of the measuring

input is replaced by one regarding the

polarization current (-127...127 uA),

or the

polarization voltage (-1270...1270 mV, in steps of 10

mV)

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on the

switchover from the inactive basic status to a

measurement. "OFF" means the test is not performed.

Temperature (-170.0...500.0 °C)

If a T sensor is connected, the temperature is

measured continuously and the pH values are

corrected accordingly.

The last measured value is entered as the

"temperature" parameter.

>stop conditions

stop V: abs.
**titr.

stop V 99.99 ml
**titr.

factor 999999
**titr.

filling rate max. ml/min
**titr.

Stop conditions

Type of stop volume (abs., rel., OFF)

"abs": absolute stop volume in mL

"rel.":

relative stop volume to sample size.

"OFF": stop volume switched off, not monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...±999 999)

Calculated as: Stop V in mL = factor

*

sample size

Filling rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min