Overview

751 GPD Titrino

3

Rear view of instrument:

7

RS232 interfaces
2 separate interfaces for the connection of printer, balance, and com-
puter

8

Remote lines (input/output)
for the connection of the Remote Box, Sample Changers, robots etc.

9

Connection of electrodes and temperature sensor
2 high-impedance measuring inputs for pH and U measurements.

They can either be used separately or for differential potentiometry,
see page 215.
Important: If you work with both measuring inputs in the same ves-
sel, the same reference electrode must be used.

1 measuring input for polarized electrodes, e.g. KF electrode

1 measuring input for PT100 or Pt1000 temperature sensor

10

Connection for separate keypad

11

Connection for power cable
With power supplies where the voltage is subject to severe HF distur-
bances, the Titrino should be operated via an additional power filter,
e.g. Metrohm 615 model.

12

Connection for external dosing units D1 and D2
685 Dosimats or 700 Dosinos

13

Connection for stirrer
728 Magnetic Stirrer, 722 Rod Stirrer, 703 or 727 Ti Stand
Supply voltage: 10 VDC (I

200 mA)

14

Rating plate
with fabrication, series and instrument number