Metrohm 751 GPD Titrino User Manual

Page 188

Advertising
background image

3.3 Properties of the RS 232 Interface

751 GPD Titrino

185

3.3.3 What can you do if the data transfer does not work?

Problem

Questions for remedial action

No characters can be received on

a connected printer.

- Are the instruments switched on and cables plugged in

correctly?

- Is the printer set to ”on-line”?

- Are baud rate, data bit and parity the same on both

instruments?

- Is the handshake set properly?

If everything seems to be ok, try to print a report with the key

sequence <PRINT><SMPL DATA><ENTER>. If this report

is printed out correctly, check if reports are defined in key

<DEF>.

No data transmission and the

display of the Titrino shows an

error message.

- error 42: Transmission error. Is the printer set to "on-line"?

Is the connection cable properly wired?

- error 43: Data output of the Titrino disabled for longer than 6

s by XOFF.

- error 36-39: Receive error. Are the RS settings the same on

both devices?

The received characters are

garbled.

- Are the RS settings the same on both devices?

- Has the correct printer been selected?

- Data transfer has been interrupted on the hardware side

during the printout of a curve. Re-establish connections and

switch printer off/on.

Wrong line spacing.

The printer does not emulate completely the preset mode.

Usually these problems arise with the IBM mode. Set the printer

to a different mode (e.g. Epson).

Printout of titration curve is not

ok. Other reports are printed ok.

Handshake is necessary for the printout of curves.

- Is your cable correctly wired? (The DTR of the printer has to

be connected to the CTS of the Titrino.)

- Set "HWs" for the handshake of the Titrino. Configure the

printer such that its DTR is set (possibly with DIP switches).

Advertising