Metrohm 751 GPD Titrino User Manual

Page 66

Advertising
background image

2.6 Parameters, key <PARAM>

751 GPD Titrino

63

max.rate max. ml/min
**titr.

min.rate 5.0 ul/min
**titr.

direction: auto

start V: OFF

start V 0.0 ml

factor 0

dos.rate max. ml/min
**titr.

pause 0 s

time interval 10 s

Maximum rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

This parameter primarily determines the addition rate

outside the control range, see also page 68.

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Minimum rate (0.01...9999 uL/min)

This parameter determines the rate within the control

range.

Direction is set automatically (+, -, auto)

auto: The direction is set automatically by the Titrino

[U

init

-U

end

]).

+:

Direction of higher pH, higher voltage (more

"positive"), higher current.

- :

Direction of lower pH, lower voltage, lower

currents.

The direction is used only when "init" has been selected

as start value.

Type of start volume (OFF, abs., rel.)

"OFF":

start volume switched off

"abs":

absolute start volume in mL

"rel.":

relative start volume to sample

If "abs." is set:

Absolute start volume (0...999.99 mL)

If "rel." is set:

Factor for relative start volume (0...±999 999)

Calculated as: Start V in mL = factor

*

sample size

Dosing rate for start volume (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Waiting time (0...999 999 s)

Waiting time, e.g. for stabilization of the electrode after

start or reaction time after dosing of a start volume. The

waiting time can be aborted with <QUIT>.

Time interval (1...999 999 s)

Time interval for the entry of the measured values in the

measuring point list. The measuring point list can

contain max. 500 points.