3.2 Remote control commands

751 GPD Titrino

173

3.2.2.

3.2.2.150.

150. IInfo.As

.Assembly.B

.Bur.S

.Select internal D0

internal D0, external D1, external D2

Selection of buret for assembly functions.

3.2.2.

3.2.2.151.

151. IInfo.As

.Assembly.B

.Bur.E

.Empty

$G, $S, $H, $C

Starts the function "empty". Only possible with the burette type "Dosino". The
parameters are under the function &Config.DosPrep, see 3.2.2.118 up to
3.2.2.120.

3.2.2.

3.2.2.152.

152. IInfo.As

.Assembly.B

.Bur.P

.Prep

$G, $S, $H, $C

Starts the function "preparation". The parameters are under the function
&Config.DosPrep, see 3.2.2.110 up to 3.2.2.120.

3.2.2.

3.2.2.153.

153. AAssembly.B

.Bur.R

.Rates.F

.Forward.S

.Selected

digital, analog

analog

AAssembly.B

.Bur.R

.Rates.F

.Forward.D

.Digital

0...150, max.

max.

AAssembly.B

.Bur.R

.Rates.R

.Reverse.S

.Selected

digital, analog

analog

AAssembly.B

.Bur.R

.Rates.R

.Reverse.D

.Digital

0...150, max.

max.

Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in cur-
rent use.
Analog means rate control with the analog potentiometer on Titrino.

3.2.2.

3.2.2.154.

154. AAssembly.B

.Bur.F

.Fill

$G, $H, $C

$G starts the 'FILL' mode of the burette function.

3.2.2.

3.2.2.155.

155. AAssembly.B

.Bur.M

.ModeDis

$G, $S, $H, $C

AAssembly.B

.Bur.M

.ModeDis.S

.Selected

volume

volume, time

AAssembly.B

.Bur.M

.ModeDis.V

.V

0.0001...0.1

0.1...9999

AAssembly.B

.Bur.M

.ModeDis.T

.Time

0.25...11...86400

AAssembly.B

.Bur.M

.ModeDis.VS

.VStop

0.0001...9999, OFF

OFF

AAssembly.B

.Bur.M

.ModeDis.A

.AutoFill

ON, OFF

OFF

Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:

Limit volume for the dispensing.

.AutoFill:

ON means automatic filling after every dispensing.

3.2.2.

3.2.2.156.

156. AAssembly.M

.Meas.S

.Status

ON, OFF

OFF

AAssembly.M

.Meas.M

.MeasInput

11, 2, Diff., Ipol, Upol, Temp

AAssembly.M

.Meas.I.Ipol

±127...11...+127

AAssembly.M

.Meas.U

.Upol

±1270...400

400...+1270