Metrohm 751 GPD Titrino User Manual

Page 59

Advertising
background image

2.6. Parameters, key <PARAM>

751 GPD Titrino

56

temperature 25.0 °C

Temperature (-170.0...500.0 °C)

If a T sensor is connected, the temperature is

measured continuously and the pH values are

corrected accordingly.

The last measured value is entered as the

"temperature" parameter.

Without a T sensor, the temperature entered manually

applies.

>stop conditions

stop V: OFF
**titr.

stop V 99.99 ml
**titr.

factor 999999
**titr.

filling rate max. ml/min
**titr.

Stop conditions

If a stop is not made after the dispensing time or after

the dispensed volume.

Type of stop volume (abs., rel., OFF)

"abs":

absolute stop volume in mL

"rel.":

relative stop volume to sample

"OFF":

stop volume switched off, not

monitored.

If "abs." is set:

Absolute stop volume (0...99 999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...± 999 999)

Calculated as: Stop V in mL = factor * sample size

Filling rate (0.01...150 mL/min, max.)

<CLEAR> sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

>monitoring

meas.mode: OFF
**titr.

meas.input: 1

Monitoring of measured values

Limit value violations are marked in the measuring

point list.

Measured values and temperatures are only stored in

the measuring point list if their monitoring is active.

Measured values (pH, U, Ipol, Upol, OFF)

If

"OFF" follow the requests:

Measuring input (1, 2, diff.)

Request only with measuring modes pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 215.