3.2. Remote control commands

751 GPD Titrino

150

Parameter for SET, KFT: Pause time in s. Runs before dosing the start volume.

3.2.2.

3.2.2.34.

34. M

Mode.P

.Parameter.T

.TitrPara.E

.ExtrT

00...999 999

Parameter for SET, KFT: Extraction time in s.

3.2.2.

3.2.2.35.

35. M

Mode.P

.Parameter.T

.TitrPara.TD

.TDelta

1...22...999 999

Parameter for SET, KFT, STAT: Time interval in s for the entry of a measure-
ment point in the list of measured points.

3.2.2.

3.2.2.36.

36. M

Mode.P

.Parameter.P

.Presel.C

.Cond

ON, OFF

OFF

M

Mode.P

.Parameter.P

.Presel.DDriftDisp

ON

ON, OFF

M

Mode.P

.Parameter.P

.Presel.DC

DCor.T

.Type

auto, man., OFF

OFF

M

Mode.P

.Parameter.P

.Presel.DC

DCor.V

.Value

0.0

0.0...99.9

Parameters for SET, KFT:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.

3.2.2.

3.2.2.37.

37. M

Mode.P

.Parameter.C

.CtrlPara.E

.EP

pH:

0...

±

20.00, OFF

OFF

U, Ipol: 0...

±

2000, OFF

OFF

Upol: 0...

±

200.0, OFF

OFF

M

Mode.P

.Parameter.C

.CtrlPara.U

.UnitEp

read only

Parameters for KFT, STAT: Setting of the EP, resp. Control point as a pH value,
in mV (with U and Ipol) or uA (with Upol). The corresponding unit can be read
with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 uA.

3.2.2.

3.2.2.38.

38. M

Mode.P

.Parameter.C

.CtrlPara.D

.Dyn

pH: 0.01...20.00, OFF

OFF

U, Ipol: 1...2000, OFF

OFF

Upol: 0.1...200.0, OFF

OFF

M

Mode.P

.Parameter.C

.CtrlPara.UnitD

.UnitDyn

read only

M

Mode.P

.Parameter.C

.CtrlPara.M

.MaxRate

0.01...10

10...150, max.

M

Mode.P

.Parameter.C

.CtrlPara.Mi

.MinIncr

0.1...9.9, min.

min.

Parameters for KFT, STAT: Control parameters.
.Dyn: Dynamics (control range) in pH, mV (with U and Ipol) or uA (with Upol).

The corresponding unit can be read with .UnitDyn. Standard with
KFT Ipol 100, with KFT Upol 10.

.MaxRate: Maximum allowed titration rate in ml/min. Max. means maximum

possible rate with mounted Exchange Unit.

.MinIncr: Parameter for KFT. Minimum volume increment in ul.

3.2.2.

3.2.2.39.

39. M

Mode.P

.Parameter.P

.Presel.O

.Oven

COM1, COM2, no

no

Parameter for KFT.