3.2 Remote control commands

751 GPD Titrino

167

scalc srt: Short report of the silo calculations.
prep:

Preparation report.

calc:

Calculation report of the current method.

act dir:

Methods of the current directory of the memory card.

mem card: All methods of the memory card.
all:

All reports.

ff:

Form feed on printer.

Reports which are sent from the Titrino are marked with space (ASCII 32) and '
at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.

3.2.2.

3.2.2.131.

131. IInfo.C

.CalibrationData

$G

IInfo.C

.CalibrationData.I.Inp1.p

.pHas

-20.00...7.00

7.00...+20.00

IInfo.C

.CalibrationData.I.Inp1.S

.Slope

-9.999...1.000

1.000...+9.999

IInfo.C

.CalibrationData.I.Inp1.T

.Temp

-170.0...25.0

25.0...+500.0

IInfo.C

.CalibrationData.I.Inp1.D

.Date

read only

IInfo.C

.CalibrationData.I.Inp1.E

.ElectrodeId

read only

identical for .Inp2 and .Diff

pH calibration data for measuring input 1. After the calibration, the data are en-
tered automatically together with the date of the calibration and the electrode
identification, see 3.2.2.71.
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.

3.2.2.

3.2.2.132.

132. IInfo.P

.PrepData.DD0.DDate

read only

IInfo.P

.PrepData.DD0.TTime

read only

IInfo.P

.PrepData.D1

D1.TType

read only

IInfo.P

.PrepData.D1

D1.DDate

read only

IInfo.P

.PrepData.D1

D1.TTime

read only

identical to .D2

Data from the last correctly executed preparation.

3.2.2.

3.2.2.133.

133. IInfo.Ch

.Checksums

$G

IInfo.Ch

.Checksums.M

.MPList

read only

IInfo.Ch

.Checksums.A

.ActualMethod

read only

The checksums can be used to identify the content of a file unequivocally, e.g.
files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList:

Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the
same results of the checksum.

3.2.2.

3.2.2.134.

134. IInfo.D

.DetermData

$G

IInfo.D

.DetermData.W

.Write

ON, OFF

OFF

IInfo.D

.DetermData.M

.MPList.1.A

.1.Attribute

read only/read+write