4.1. Error and special messages

751 GPD Titrino

190

rate too low

With DOS the calculated rate is too low. It can not be achieved

with the mounted Exchange Unit.

Exit: <CLEAR> and use Exchange Unit with smaller buret

volume or change dosing conditions.

rates not corresponding

With STAT the number of rates that could be calculated is not

equal to the number of time windows set or the mean rate C80

could not be calculated. The calculation of a rate requires at

least 4 points.

result out of limits

The result lies outside the limits which were defined in the
method, see page 75.
Exit: Calculate result again or new start.

same buffer

In the calibration sequence the voltage difference between the

first and second buffer is < 6 mV.

Exit: <QUIT> and change buffer or <STOP> (abort

calibration).

sample size out

The sample size is outside the limits which are defined in the
method, see page 19.
Exit: Enter new sample size.

save lines OFF

The function "save lines" is not active although a submethod of

TIP contains assignments to C24 or C25.

Exit: <CLEAR> and switch on "save lines" under <SMPL

DATA> key.

Warning: The results of this line will not be saved.

second TIP call

In TIP no further TIP can be called up as a submethod.

Exit: <CLEAR> and define new sequence.

service is due

The service interval has elapsed. Contact Metrohm service so
that the Titrino can be serviced. This message will appear
each time the Titrino is switched on.
Exit: New start.

silo empty

The silo memory is switched in but empty and a titration has

been started. Corrective action: At least the first 1 silo line

before starting the first titration.

Exit: <CLEAR>.

silo full

The silo memory is full up. Corrective action: If you have filled

less than 255 silo lines, you can create more space by

deleting old methods no longer needed. 1 silo line needs

18...120 bytes.

Exit: <CLEAR>.

stop EP reached

A DET or MET titration was stopped as the stop criterion "stop

EP" was reached.

stop meas.val.reached

A DET or MET titration was stopped as the stop measuring

value pH, U or I was reached.

stop rate reached

STAT has been stopped as the stop rate has been reached.

stop time reached

SET or STAT has been stopped as the stop time has been

reached.

stop V reached

The determination has been stopped as the stop volume has

been reached.

system error 3

The instrument adjustment data have been overwritten.

Exit: <CLEAR>. Default adjustment data are set. The error

message appears each time the instrument is switched on

until it has been readjusted (Metrohm service).