5.1. Setting up and connecting the instruments
751 GPD Titrino
210
5.1.5 Connection of a Sample Changer
With 6.2141.030 cable (instead of 6.2141.020), two Titrinos can be
connected to the 730 or 760 Sample Changer at the same time.
•
The "Remote" socket allows not only connection of a sample
changer but also additional control functions. Pin assignment of
the "Remote" socket and control possibilities, see page 223.
•
If a calibration has to be performed with the sample changer, the
calibration parameter "sample changer:" must be set to "ON".
•
In connections with the sample changer, "auto start" should be
set to "OFF" in the <CONFIG> key. The start command is given
by the Sample Changer.
C
B
A
D
751
6.2141.020 Cable
730/760