3.2. Remote control commands

751 GPD Titrino

176

3.2.2.

3.2.2.164.

164. Se

Setup.M

Mode.S

.StartWait

ON, OFF

OFF

Se

Setup.M

Mode.F

.FinWait

ON, OFF

OFF

Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP

(holding point after AutoInfo !".T.GC").

.FinWait:

Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").

3.2.2.

3.2.2.165.

165. Se

Setup.S

.SendMeas.S

.SendStatus

ON, OFF

OFF

Se

Setup.S

.SendMeas.I.Interval

0.08...44...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

3.2.2.167 and 3.2.2.168) in the inputted interval is active.

.Interval:

Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.167 and 3.2.2.168.
The inputted value is rounded off to a multiple of 0.08. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the interface
and on the type of device connection. With "MPList" the measured
values are sent at the time of their entry into the measured point list.

The automatic transmission is switched on/off with 'SendStatus'.

3.2.2.

3.2.2.166.

166. Se

Setup.S

.SendMeas.Sel

.Select

Assembly, Titrator

Titrator

Selection of the unit of which the measured values should be sent (3.2.2.167 or
3.2.2.168).

3.2.2.

3.2.2.167.

167. Se

Setup.S

.SendMeas.A

.Assembly.C

.CyclNo

ON, OFF

OFF

Se

Setup.S

.SendMeas.A

.Assembly.V

.V

ON, OFF

OFF

Se

Setup.S

.SendMeas.A

.Assembly.M

.Meas

ON, OFF

OFF

Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.165):
.CyclNo:

Cycle number of the potential measurement. Together with the cy-
cle time (3.2.2.149), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it
is always incremented as long as the instrument remains switched
on.

.V:

Volume

.Meas:

Measured value associated to the cycle number.

The unit "assembly" must be preset (see 3.2.2.166).

3.2.2.

3.2.2.168.

168. Se

Setup.S

.SendMeas.T

.Titrator.C

.CyclNo

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.V

.V

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.M

.Meas

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.d

.dVdt

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.dM

.dMeasdt

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.dMeasdV

.dMeasdV

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.E

.ERC

ON, OFF

OFF

Se

Setup.S

.SendMeas.T

.Titrator.T

.T

ON, OFF

OFF