4.2. Diagnosis

751 GPD Titrino

203

11 External dosing units D 1 or D 2

This test makes sense only when the 751 GPD Titrino is operated in connection with external

dosing units (685 Dosimat or 700 Dosino).
The spindle drive and the cock changeover drive of the internal dispensing unit can be checked
with the aid of the function <prep> (see page 99).

Connect the external dosint units D 1 or D 2 respectively.

Insert exchange unit and put the burette tip into a collecting receptacle.

Set the parameters of this function accord. to the following list (for setting the parameters

see configuration >prep.dosing elements).

Parameters for 700 Dosino

Parameters for 685 Dosimat

configuration

configuration

>prep.dosing elements

>prep.dosing elements

power ON prep: OFF

power ON prep: OFF

report: OFF

report: OFF

dos.element: external D1

or

D2

dos.element: external D1

or

D2

warn.interv.DX OFF min

warn.interv.DX OFF min

dos.drive DX: Dosino

dos.drive DX: Dosimat

outlet: tip

volume DX 3.5 ml

len.dos.tub.DX 40.0 cm

cycles DX 2

diam.dos.tub.DX 2.0 mm

dos.rate DX max. ml/min

ml/min

len.asp.tub.DX. 25.0 cm

fill rate DX max. ml/min

ml/min

diam.asp.tub.DX 2.0 mm

dos.rate DX max. ml/min

fill rate DX max. ml/min

Press <prep> several times, until

extern D1 prep

or

extern D2 prep

Press <START>

The external dosing unit executes the following jobs:

a) To expell the total volume.
b) to fill the filling tube
c) to fill the cylinder
d) to fill the dosing tube.

12 Setting up original arrangement

Reconnect all peripherals disconnected at the start of the diagnostic routine and perform a

short function test with these.