Metrohm 751 GPD Titrino User Manual

Page 132

Advertising
background image

3.2 Remote control commands

751 GPD Titrino

129

*Parameter Tree part "Parameters for TIP"

Ã

.

Sequence

Sequence

³ Ã

.1

Step 1

³ ³ Ã

.Select

Step selection

method,pause,L4 output,L6 output,

³ ³

info,prep,stirrer,OFF

3.2.2.74.

³ ³ Ã

.Method

Method from mem.or card

special

3.2.2.75.

³ ³ Ã

.Pause

Waiting time

0...999 999, INF

ditto

³ ³ Ã

.L4Output

Line L4

active,inactive,pulse,OFF

ditto

³ ³ Ã

.L6Output

Line L6

active,inactive,pulse,OFF

ditto

³ ³ Ã

.Info

Display information

up to 16 ASCII char.

ditto

³ ³ Ã

.Prep

Preparation of titrating buret

internal D0, external D1,

³ ³ ³

external D2

ditto

³ ³ Ã

.Stirrer

Stirrer

ON, OFF

ditto

³ Ã

:

up to 30 steps

³ ³

Ã

.

Statistics Statistics

³ Ã

.Status

Status of statistics calculation

ON, OFF

3.2.2.24.

³ Ã

.MeanN

No. of individual determinations

2...20

ditto

³ Ã

.ResTab

Result table

³ ³ Ã

.Select

original,delete n,delete all

ditto

³ ³ Ã

.DelN

Deletion of individual results

1...20

ditto

³ ³

Ã

.

Presel

Preselections

³ Ã

.IReq

Request of Id's after start

id1, id1&2, all, OFF

3.2.2.26.

³ Ã

.SReq

Request of sample size after start

value, unit, all, OFF

ditto

³ Ã

.LimSmplSize

Limits for sample size

3.2.2.27

³ ³ Ã

.Status

Status of limit control

ON, OFF

ditto

³ ³ Ã

.LoLim

Lower limit

0.0...999 999

ditto

³ ³ Ã

.UpLim

Upper limit

0.0...999 999

ditto

³ Ã

.MeasMode

Measuring mode for man.meas.

pH,U,Ipol,Upol,T,OFF

3.2.2.76.

³ Ã

.MeasInput

Measuring input

1, 2, diff.

ditto

³ Ã

.Ipol

Polarization current

0...1...

±

127

ditto

³ Ã

.Upol

Polarization voltage

0...400...

±

1270

ditto

³ Ã

.PolElectrTest

Test for polarized electrodes

ON, OFF

ditto

³ Ã

.Temp Titration temperature

-170.0...25.0...500.0

ditto