<7. Startup>

7-12

IM 11M12A01-02E

7.11 Checking Contact I/O

Conduct the contact input and output checking as well as operational checking of the solenoid

valves for automatic calibration.

7.11.1 Checking Contact Output

To check the contact output, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.

Select “Maintenance” in that display.

(2) Select “Contact check” then “Output contacts” in the Maintenance panel display to display

the Output contacts display (see Figure 7.15.2).

(3) In this display, select the desired output contact for checking. The display, which enables

the closing and opening of contacts, then appears. Use the display to conduct continuity

checking.

Output contact 1 : Open

Output contact 2 : Open

Output contact 3 : Open

Output contact 4 : Open

Enter

Output contacts

F7.15.2E.ai

Figure 7.15.2 Output Contact Check Panel Display

NOTE

• If you conduct an open-close check for contact output 4, Error 1 or Error 2 will occur.

This is because the built-in heater power of the detector, which is connected to contact

output 4, is turned off during the above check. So, if the above error occurs, reset the

equipment or turn the power off and then back on to restart (refer to Section 10.4, Reset,

later in this manual).