<10. Other Functions>

10-4

IM 11M12A01-02E

10.1.1.11 Maximum Oxygen Concentration

The maximum oxygen concentration and the time of its occurrence during the period specified

in the Averaging display are displayed. After the present monitoring interval has elapsed, the

maximum oxygen concentration that has been displayed so far will be cleared and a new

maximum oxygen concentration will be displayed. If the setup period of time is changed, the

current maximum oxygen concentration will be displayed (for more details, see Section 8.6.2

earlier in this manual).

10.1.1.12 Minimum Oxygen Concentration

The minimum oxygen concentration and the time of its occurrence during the period specified

in the Averaging display are displayed. If the setup period elapses, the minimum oxygen

concentration that has been displayed so far will be cleared and a new minimum oxygen

concentration will be displayed. If the setup period of time is changed, the current minimum

oxygen concentration will be displayed (for more details, see Section 8.6.2 earlier in this manual).

10.1.1.13 Average Oxygen Concentration

The average oxygen concentration during the periods over which average values are calculated

is displayed. If the setup period elapses, the average oxygen concentration that has been

displayed so far will be cleared and a new average oxygen concentration will be displayed. If the

setup period of time is changed, the current average oxygen concentration will be displayed (for

more details, see Section 8.6.2 earlier in this manual).

10.1.1.14 Heater On-Time Ratio

The probe sensor is heated to and maintained at 750 °C. When the sample gas temperature is

high, the amount of heater ON-time decreases.

10.1.1.15 Time

The current date and time are displayed. These are backed up with built-in batteries, so the clock

continue the run even if the power is switched off.

10.1.1.1.16 History of Calibration Time

The calibration-conducted dates and times, and span gas and zero gas ratios for the past ten

calibrations are stored in memory.

10.1.1.17 Power Supply Voltage

For the temperature control for the heater of the detector to work best, you should set the power

supply voltage and frequency appropriately, as the control parameters are based on this. Set the

AC supply voltage to “Low” if supply is 140 V AC or less, and to “High” if it is 180 V or more.

10.1.1.18 Power Frequency

Set the AC supply frequency setting appropriately --”Low” for 50 Hz, and “High” for 60 Hz.