<3. Installation>

3-9

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

<Wall Mounting>
(1) Drill mounting holes through the wall as shown in Figure 3.8.

190

126.5

Four holes 6 mm in diameter

for M5 screws

Unit: mm

F3-8E.ai

Figure 3.8 Mounting holes

(2) Mount the converter. Secure the converter on the wall using four screws.

Note: For wall mounting, the bracket and bolts are not used.

F3-9E.ai

Figure 3.9 Wall Mounting

<Panel Mounting>

(1) Cut out the panel according to Figure 3.10.

274+2

0

183

+2

0

F3-10E.ai

Unit: mm

Clamp screw

Washer

Bolt

Fitting

F3-11E.ai

Figure 3.10 Panel cutout sizes

Figure 3.11 Panel mounting

(2) Remove the fitting from the converter by loosening the four screws.
(3) Insert the converter case into the cutout hole of the panel.
(4) Attach the mounting fitting which is once removed in step (2) again to the converter.
(5) Firmly fix the converter to the panel. Fully tighten the two clamp screws to hold the panel with

the fitting.