<2. Specifications>

2-21

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

2.7

Other Equipments

2.7.1

Dust Filter for the Detector (K9471UA)

This filter is used to protect the detector sensor from a corrosive dust components or high velocity

dust in recovery boilers and cement kilns.
Sample gas flow rate is needed to be 1 m/sec or more to replace gas inside zirconia sensor.

• Standard specification
Applicable detector: Standard-type detector for general use (the sample gas flow should be

approximately perpendicular to the probe.)

Mesh:

30 microns

Material:

Carborundum (Filter), SUS316 (JIS)

Weight:

Approx. 0.2 kg

Part No.

Description

K9471UA

Filter

K9471UX

Tool

Ø

51

32

10

Carborundum filter (SiC)

Detector

Screw

Increasing of insertion length

F3-2E.ai

Attach the filter unit to the tip of

the detector by screwing it clockwise.

2.7.2

Dust Guard Protector (K9471UC)

Recommended to be used when sample gas is likely to flow directly into the cell due to its flow

direction in the stack or the like, flammable dust may go into the cell, or water drops are likely to

fall and remain in the cell during downtime or the like due to the installation position.
Material:

SUS316

Weight:

Approx. 0.3 kg

4-

Ш

6

122

Ш

50.8

100

Increasing of insertion length

F11-1E.ai