<10. Other Functions>

10-10

IM 11M12A01-02E

NOTE

• If the blow back is executed with an input contact, it must be preset in the Input contact

settings (for more details, see Section 8.5, earlier in this manual).

• In Section 8.4, Output Contact Setup, earlier in this manual, set the contact used as the

blow back switch beforehand.

• Do not set any other function for the contact used as the blow back switch.

Otherwise, blow back may be activated when the contact is closed by any other function.

• No blow back is executed during calibration or maintenance service. If automatic blow back

reaches the preset start time during calibration or maintenance service, blow back will be

executed after completing the calibration or maintenance service and after the equipment

returns to the measurement mode.

• If automatic blow back reaches the preset start time during semi-automatic blow back, the

current automatic blow back will not be executed.

• If you set the blow back interval at 000 days, 00 hours, only the first blow back is then

executed. No subsequent blow backs will be executed.

• If a past date is set for the Start time, no blow back will be executed.

10.2.1.5 Default Setting

When the analyzer is delivered, or if data are initialized, the blow back settings are by default, as

shown in Table 10.4.

Table 10.4 Blow back Default Setting

Item

Default setting

Mode

No function (“invalid”)

Hold time

10 minutes, 00 seconds

Blow back time

10 minutes, 00 seconds

Interval

30 days, 00 hours

Start date

01/ 01/00

Start time

0:00