10 checking current loop, Checking current loop -11, Ma - output loop check – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 99

Advertising
background image

<7. Startup>

7-11

IM 11M12A01-02E

7.10 Checking Current Loop

The set current can be output as an analog output.
(1) Press the [Setup] key on the Basic panel display to display the Execution/Setup display.

Then select “Maintenance” in the Execution/Setup display.

(2) Select “mA-output loop check” in the Maintenance panel display to display the mA - output

loop check display, enabling you to check “mA-output1” and “mA-output2”.

Select the desired output terminal for current-loop checking (see Figure 7.15.1).

(3) At the time of entering the numeric-data entry display, the output current will change to 4 mA

(default value). If the desired current is entered, the corresponding output will be provided.

mA - output 1

mA - output 2

Enter

mA - output loop check

F7.15.1E.ai

Figure 7.15.1 “mA - output loop check” Panel Display

8th Edition : Jan.13,2012-00

Advertising
This manual is related to the following products: