<3. Installation>

3-4

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

<Detector with a probe protector (Model ZO21R-L-- *B for enhance forth>
The detector is used with a probe protector to support the probe (ZR22G) when the probe length

is 2.5 m or more and it is mounted horizontally.
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the

probe insertion hole.
(2) Make sure that the sensor assembly mounting screws (four bolts) at the tip of the detector are

not loose.

(3) Mount the detector so that the reference gas and calibration gas inlet faces downward.

1050,1550,2050

Notch

Unit: mm

F3-3E.ai

Gasket (t1.5)

Reference gas inlet

Calibration gas inlet

Flow direction of the

sample gas

Φ60.5

Probe top

Figure 3.3 Probe protector (supporting the mounting strength)

<Detector with a probe protector (Model ZO21R-L-- *B for dust wear protect>
The detector is used with a probe protector to prevent the sensor from being worn by dust

particles when there is a high concentration of dust and gas flow exceeds 10 m/sec (pulverized

coal boiler or fluidized-bed furnace).
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the

probe insertion hole. The probe protector should be installed so that the notch is downstream

of the sample gas flow.
(2) Make sure that the sensor assembly mounting screws (four bolts) at the probe tip are not

loose.

(3) Where the detector is mounted horizontally, the reference gas and calibration gas inlet should

face downward.

NOTE

When the probe protector is used in the ZR22G with pressure compensation (-P), instrument air

leaking from the probe protector may affect the measured value.

Direction of the sample gas flow

Mount the protector so that the notch is

the downstream side of the gas flow.

Reference gas inlet
Calibration gas inlet

Unit: mm

F3-4E.ai

Probe top

1050,1550,2050

Gasket (t1.5)

Figure 3.4 Mounting of detector with a probe protector