<2. Specifications>

2-1

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

2.

Specifications

This chapter describes the specifications for the following:
ZR22G

General use separate type detector (See Section 2.2.1)

ZO21R

Probe protector (See Section 2.2.2)

ZR22G (0.15 m)

High temperature separate type detector (See Section 2.3.1)

ZO21P

Adapter for High temperature probe (See Section 2.3.2)

ZR402G

Separate type converter (See Section 2.4)

ZA8F

Flow setting unit (See Section 2.5.1)

ZR40H

Automatic calibration unit (See Section 2.5.2)

ZO21S

Standard gas unit (See Section 2.6)

2.1

General Specifications

2.1.1

Standard Specifications

Measured Object:

Oxygen concentration in combustion exhaust gas and mixed gas

(excluding inflammable gases, may not be applicable corrosive gas such

as ammonia is present-check with YOKOGAWA)

The sampling gases containing a corrosive gas such as ammonia or

chlorine may be applicable to our oxygen gas analyzer. In this case,

contact with YOKOGAWA and its agency.

Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O

2

Output Signal:

4 to 20 mA DC (maximum load resistance 550 Ω)

Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O

2

(in 1 vol% O

2

), or partial range

Digital Communication (HART): 250 to 550 Ω, depending on number of field devices connected

to the loop (multi-drop mode).

Note: HART is a registered trademark of the HART Communication Foundation.

Display Range:

0 to 100 vol% O

2

Warm-up Time:

Approx. 20 min.

Repeatability:

(Excluding the case where the reference gas is by natural convection)

±0.5% Maximum value of set range. (range from 0 to 5 vol% O

2

or more

and less than 0 to 25 vol% O

2

range)

±1% Maximum value of set range.

(range from 0 to 25 vol% O

2

or more and up to 0 to 100 vol% O

2

range)

Linearity:

(Excluding standard gas tolerance)

(Excluding the case where the reference gas is by natural convection)

(Use oxygen of known concentration (within the measuring range) as the

zero and span calibration gases.)

±1% Maximum value of set range ;

0 to 5 vol% O

2

or more and less than 0 to 25 vol% O

2

range

(Sample gas pressure: within ±4.9 kPa)

±3% Maximum value of set range ;

0 to 25 vol% O

2

or more and less than 0 to 50 vol% O

2

range