<5. Wiring>

5-2

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

CAUTION

• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected in ways equivalent to JIS D style (Class 3)grounding

(the grounding resistance is 100 Ω or less).
• Special consideration of cable length should be taken for the HART communication, For the

detail, refer to Section 1.1.2 of the IM 11M12A01-51E “ Communication Line Requirements “.

5.1.1

Terminals for the External Wiring in the Converter

Open the front door and remove the terminal covering plate to gain access to the converter

external wiring terminals (see Figure 5.1).

CAUTION

After wiring necessary cable to the converter terminals, be sure to fix the terminal covering plate

with two screws again.

F5-2E.ai

Front door

Cable inlet

Terminals are visible

when the terminal cover is removed.

Figure 5.1 Terminals for external wiring in the converter