<8. Detailed Data Setting>

8-8

IM 11M12A01-02E

Parameter: Oxygen

Hysteresis: 0.1%O

2

Contact delay: 3 s

Setpoints

Enter

Alarms setup

F8.4E.ai

High High alarm: OFF

Set value: 1 0 0 . 0 % O

2

High alarm: ON

Set value: 1 0 0 . 0 % O

2

Low alarm: OFF

Set value: 0 . 0 % O

2

Low Low alarm: OFF

Set value: 0 . 0 % O

2

Enter

Oxygen alarms

F8.5E.ai

Figure 8.4 Alarms Setup Display

Figure 8.5 Oxygen Alarms Display

8.3.4

Default Values

When the analyzer is delivered, or if data are initialized, the default alarm set values are as

shown in Table 8.3.

Table 8.3 Alarm Setting Default Values

Item

Setting range

Default setting

Hysteresis

0 to 9.9 vol%O

2

0.1 vol%O

2

Delay time

0 to 255 seconds

3 seconds

High-high limit alarm

OFF

High-high limit alarm setpoint

0 to 100 vol%O

2

100 vol%O

2

High limit alarm

OFF

High limit alarm setpoints

0 to 100 vol%O

2

100 vol%O

2

Low limit alarm

OFF

Low limit alarm setpoint

0 to 100 vol%O

2

0 vol%O

2

Low-low limit alarm

OFF

Low-low limit alarm setpoint

0 to 100 vol%O

2

0 vol%O

2