<12. Troubleshooting>

12-4

IM 11M12A01-02E

NOTE

• Measure the thermocouple resistance value after the temperature difference between

the detector tip and the ambient atmosphere has decreased to 50 °C or less. If the

thermocouple voltage is large, accurate measurement cannot be achieved.

12.1.2.3 Error3: A/D Converter Failure/Error4: Writing-to-memory Failure

• A/D Converter Failure
It is suspected that a failure has occurred in the A/D converter mounted in the electrical circuits

inside the converter.
• Writing-to-memory Failure
It is suspected that a failure has occurred in an operation writing to the memory (EEPROM)

mounted in the electrical circuits inside the converter.

<Locating the failure, and countermeasures>
Turn off the power to the converter once and then restart the converter. If the converter

operates normally after restarting, an error might have occurred due to a temporary drop in the

voltage (falling below 85 V, the least amount of voltage required to operate the converter) or a

malfunction of the electrical circuits affected by noise. Check whether or not there is a failure in

the power supply system or whether the converter and detector are securely grounded.
If the error occurs again after restarting, a failure in the electrical circuits is suspected.

Consult the service personnel at Yokogawa.