3 piping for the reference gas, 4 piping to the high temperature probe adapter, Piping for the reference gas -7 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 67: Piping to the high temperature probe adapter -7

Advertising
background image

<4. Pipomg>

4-7

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

F4-8E.ai

Stop valve or Check valve

Piping for the calibration gas, 6 mm (O.D.)

by 4 mm (I.D.) Stainless steel pipe

Piping for the reference gas, 6 mm (O.D.)

by 4 mm (I.D.)Stainless steel pipe

Figure 4.8 Piping for the Calibration Gas Inlet

4.2.3

Piping for the Reference Gas

Reference gas piping is required between the air source (instrument air) and the flow setting unit,

and between the flow setting unit and the detector.
Insert the air set next to the flow setting unit in the piping between the air source and the flow

setting unit.
Use a 6 mm (O.D.) x 4 mm or larger (I.D.) (or nominal size 1/4 inch) stainless steel pipe between

the flow setting unit and the detector.

4.2.4

Piping to the High Temperature Probe Adapter

Refer to Section 4.1.4 for the piping.