<5. Wiring>

5-12

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.8

Contact Input Wiring

The converter can execute specified function when receiving contact signals.

To use these contact signals, wire as follows:

ZR402G

Converter

Terminal box

DI-1

DI-C

DI-2

Contact input 1

Contact input 2

F5-14E.ai

Figure 5.13 Contact Input Wiring

5.8.1

Cable Specifications

Use 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine which

cable to use.

5.8.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each cable should be terminated

corresponding to crimp-on terminals.

(2) The ON/OFF level of this contact input is identified by the resistance. Connect a contact

input that satisfies the specifications in Table 5.2.

Table 5.2 Identification of Contact Input ON/OFF

Closed

Open

Resistance

200 Ω or less

100 kΩ or more