<7. Startup>

7-14

IM 11M12A01-02E

7.12 Calibration

To calibrate this instrument, the procedure is to measure zero gas and span gas and set the

instrument to read the known concentrations. The procedure for both zero and span calibration,

or for either zero or span calibration, can be performed manually from the touch display, or can

be performed semi-automatically using contact signal inputs to start calibration, (allowing preset

calibration and stabilization times), or it can be performed automatically at preset intervals.
There are three types of calibration procedures available:
(1) Manual calibration conducting zero and span calibrations, or either of these calibrations in

turn.

(2) Semi-automatic calibration which uses the touchpanel or a contact input signal and

conducts calibration operations based on a preset calibration time and stable time.

(3) Automatic calibration conducted at preset intervals.
Manual calibration needs the ZA8F Flow Setting Unit to allow manual supply of the calibration

gases. Semi-automatic and automatic calibrations need the ZR40H Automatic Calibration Unit

to allow automatic supply of the calibration gases. The following sections set forth the manual

calibration procedures. For details on semi-automatic and automatic calibrations, consult

Chapter 9, Calibration, later in this manual.

7.12.1 Calibration Setup

7.12.1.1 Mode Setting

For the mode setting, do the following:
Press the [Setup] key in the Basic panel display to display the Execution/Setup display.

Select “Maintenance” in the Execution/Setup display to display the Maintenance panel display.

Then select “Calibration setup” to display the Calibration setup display as Figure 7.16 shows.

Select Mode in this panel, and then select “Manual”, “Semi_Auto” or “Auto”.

Enter

Mode: Manual

Points: Both

Zero gas conc: 1.00%

Span gas conc:   21.00%

Timing

Calibration setup

F7.16E.ai

Figure 7.16 Calibration Setup Display

7.12.1.2 Calibration Setting Procedures

Select “Points” (calibration procedure) in the Calibration setup display to display the “Both, Span,

Zero” selection display. In this display, select “Both.”

7.12.1.3 Calibration Gas Concentration Setting

(1) Zero gas concentration

If “Zero gas conc” is selected in the Calibration setup display, the numeric-data entry display

then appears.

Use this display to enter an oxygen concentration value for the zero gas calibration;

if the oxygen concentration is 0.98 vol%O

2

, enter 00098.