<10. Other Functions>

10-12

IM 11M12A01-02E

Table 10.5 Initialization Items and Default Values - Part 1

Item

Initialization Parameter

Default setting

Equipment

selection

Type of equipment

Not initialized

Detector

ZR22

Sample gas

Wet gas

Displayed data

Display item

1st display item

Oxygen concentration

2nd display item

Current output 1

3rd display item

Current output 2

Tag name

Deleted

Trend graph

Parameter

Oxygen concentration

Sampling interval

30 seconds

Upper limit (graph)

25 vol%O

2

Lower limit (graph)

0 vol%O

2

Automatic return time

0 min.

Language

Not initialized

Calibration data

Calibration setting

Mode

Manual

Calibration procedure

Span - zero

Zero gas concentration

1.00 vol%O

2

Span gas concentration

21.00 vol%O

2

Output hold time

10 min., 00 sec. (Semi_Auto)

Calibration time

10 min., 00 sec. (Semi_Auto)

Interval

30 days, 00 hr. (Auto)

Start date

01/01/00 (Auto)

Start time

0:00 (Auto)

Blow back

Blow back setting

Mode

No function (invalid)

(Output) hold time

10 min., 00 sec.

Blow back time

10 min., 00 sec.

Interval

30 days, 00 hr.

Start date

01/01/00

Start time

0:00

Current output

data

mA-output 1

mA-output 2

Parameter

Oxygen concentration

Min. oxygen concentration

0 vol%O

2

Max. oxygen concentration

25 vol%O

2

Output damping constant

0 sec.

Output mode

Linear

Output hold setting

Warm-up

4 mA

Set value of Warm-up

4 mA

Maintenance

Previous value held

Set value of Maintenance

4 mA

Calibration, blow back

Previous value held

Set value of Calibration, blow

back

4 mA

Error

Preset value held

Set value of Error

3.4 mA