3 installation of the zr402g converter, 1 installation location, 2 mounting of the converter – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 54: Installation of the zr402g converter -8 3.3.1, Installation location -8, Mounting of the converter -8

Advertising
background image

<3. Installation>

3-8

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

3.3

Installation of the ZR402G Converter

3.3.1

Installation Location

The following should be taken into consideration when installing the converter:
(1) Readability of the indicated values of oxygen concentration or messages on the converter

display. Easy and safe access to the converter for operating keys on the panel.
(2) Easy and safe access to the converter for checking and maintenance work.
(3) An ambient temperature of not more than 55 °C and little change in temperature

(recommended within 15 °C in a day).
(4) The normal ambient humidity (recommended between 40 to 75% RH) and without any

corrosive gases.
(5) No vibration.
(6) Near to the detector.
(7) Not in direct rays of the sun. If the sun shines on the converter, prepare the hood (/ H) or other

appropriate sunshade.

3.3.2

Mounting of the Converter

The converter can be mounted on a pipe (nominal JIS 50A: O.D. 60.5 mm), a wall or a panel. The

converter can be mounted at an angle to the vertical, however, it is recommended to mount it on

a vertical plane.
Mount the converter as follows.
<

Pipe Mounting>

(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting

the converter. (Converter weighs approximately 6 kg.)
(2) Mount the converter on the pipe. Fix it firmly on the pipe in the procedure described in Figure

3.7.

A pipe to be mounted (nominal JIS 50A)

Bracket

Nut

Washer

Bolt

*Bolt

*Washer

*Fitting

* Note: These fittings are attached to the converter when it is delivered.

F3-7E.ai

Mounting procedure

1. Put four bolts in the holes on the fitting.
2. Clamp the pipe with the fitting and a bracket,
with the four bolts passing through the bracket holes.
3. Secure the fitting and the bracket tightly to the pipe with
four washers and nuts.

Figure 3.7 Pipe Mounting