<11. Inspection and Maintenance>

11-3

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

(5) Clean the sensor assembly, especially the metal O-ring contact surface to remove any

contaminants adhering to that part. If you can use any of the parts from among those

removed, also clean them up to remove any contaminants adhering to them.

(Once the metal O-ring has been used, it can not be reused. So, be sure to replace it.)

3. Part assembly procedure
(1) First, install the contact. Being careful not to cause irregularities in the pitch of the coil spirals

(i.e., not to bend the coil out of shape), place it in the ringed groove properly so that it forms

a solid contact.

Groove in which the contact

(E7042BS) is placed

F11.2E.ai

Figure 11.2 Installing the Contact

(2) Next, make sure that the O-ring groove on the flange surface of the sensor is clean.

Install the metal O-ring in that O-ring groove, and then insert the sensor in the probe while

turning it clockwise. After inserting it until the metal O-ring comes in contact with the probe’s

O-ring contact surface, properly align the U-shaped pipe insertion holes with the bolt

openings.

(3) Attach the U-shaped pipe to its support with filter, then fully insert the U-shaped pipe and its

support into the probe.

(4) Coat the threads of the four bolts with anti-seize grease and then screw them in along with

the washers. First, tighten the four bolts uniformly by hand, and then use a torque wrench

to tighten all areas of the metal O-ring uniformly, that is, to make sure the sensor flange is

perfectly horizontal to the O-ring’s working face in the probe.

This is done by tightening first one bolt and then its opposing bolt each 1/8 turn, and then

one of the other bolts followed by its opposing bolt, each also 1/8 turn. This continues

in rotating fashion until they are all fully tightened with the torque wrench preset to

approximately 5.9 N • m. If they are not uniformly tightened, the sensor or heater may be

damaged.

Replacement of the sensor assembly is now complete. Install the detector and restart

operation. Calibrate the instrument before making a measurement.