<5. Wiring>

5-9

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.4

Wiring for Analog Output

This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.

Maintain the load resistance including the wiring resistance at 550 Ω or less.

Shielded cable

Receiver 1

Receiver 2

ZR402G

Converter

1
2

1
2

AO-1(+)
AO-1(-)
AO-2(+)
AO-2(-)

FG

F5-9E.ai

Figure 5.8 Wiring for analog output

5.4.1

Cable Specifications

For this wiring, use a 2-core or a 4-core shielded cable.

5.4.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each wire in the cable should be

terminated corresponding to crimp-on terminals. Ensure that the cable shield is connected

to the FG terminal of the converter.

(2) Be sure to connect “+” and “-” polarities correctly.

5.5

Power and Grounding Wiring

This wiring supplies power to the converter and grounds the converter/detector.

ZR402G

Converter

Jumper plate

L N G

ZR22G

Detector

Ground

100 - 240 V AC

50/60 Hz

Grounding to the ground terminal

on the converter case

Converter case

FG terminal Lock washer

Crimp-on terminal of

the ground wire

F5-10E.ai

Figure 5.9 Power and Grounding wiring

5.5.1

Power Wiring

Connect the power wiring to the L and N terminals of the converter. Proceed as follows:
(1) Use a 2-core or a 3-core cable.
(2) The size of converter terminal screw threads is M4. Each cable should be terminated

corresponding to crimp-on terminals.