10.3 Parameter Initialization

Parameter settings can be initialized to the factory default settings. Initialization can be done

for all parameters or for individual parameters. The parameters that can be initialized and their

defaults are listed in Table 10.5.
<Initialization procedure>

(1) On the Basic panel display, touch the [Setup] key to display the Execution/Setup display.
(2) Select “Commissioning”, next “Others” then “Defaults”. A display like Figure 10.11. appears.
(3) Select the desired item to be initialized then a display like Figure 10.12 appears.

Select “Defaults start” then initialization starts.

Display

Calibration

Blow back

mA-outputs

Alarms

Contacts

Others

All data

Enter

Defaults

F10.11E.ai

Enter

Defaults start

Cancel

Defaults

Display

Figure 10.11 Defaults Display

Figure 10.12 Initialization Start Display

CAUTION

• Do not turn off the power during initialization. Otherwise, initialization will not be performed

properly.