<7. Startup>

7-17

IM 11M12A01-02E

(7) Follow the instructions in the display as in Figure 7.23 to turn on the zero gas flow. To do

this, open the zero gas flow valve for the Flow Setting Unit and adjust that valve to obtain a

flow of 600 ± 60 ml/min. The valve should be adjusted by loosening its lock nut and slowly

turning the valve shaft counterclockwise. Use the calibration gas flowmeter to check the

flow.

Enter

Open zero gas valve.
Set flow zero gas to
600ml/min.

Valve opened

Cancel calibration

Manual calibration

F7.23E.ai

Enter

21.00%

0.5min./div

1.00%

21.0%O

2

CAL. TIME

Manual calibration

F7.25E.ai

Figure 7.23 Zero Gas Flow Display

Figure 7.24 Zero Gas, Trend Graph

(for Manual calibration)

(8) If “Valve opened” is selected as in Figure 7.23 , an oxygen concentration trend graph (with

the oxygen concentration being measured) appears (see Figure 7.24). The CAL.TIME in

the bottom area of the panel flashes. Observe the trend graph and wait until the measured

value stabilizes in the vicinity of the zero gas concentration on the graph. At this point, no

calibration has been executed yet, so even if the measured value is above or below the zero

gas concentration value, no problem occurs.

(9) After the measured value has stabilized, press the [Enter] key to display the “Zero

calibration complete” display shown in Figure 7.25. At this point, the measured value is

corrected to equal the zero gas concentration setting. Close the zero gas flow valve. The

valve lock nut should be tightened completely so that the zero gas does not leak.

Enter

Zero calibration
Close the zero gas valve.

Span calibration

End

Manual calibration

F7.26E.ai

Figure 7.25 Zero Calibration Complete Display

(10) Select “End” in the display as shown in Figure 7.25. An oxygen concentration trend graph

(with the oxygen concentration being measured) appears and HOLD TIME then flashes.

This time is referred to as the output-stabilization time. If the HOLD TIME has been set

with the output hold setting, the analog output remains held (refer to Section 8.2, Output

Hold Setting, later in this manual). Manual calibration is completed when the preset hold

(output stabilization) time elapses. This hold (output stabilization) time is set to 10 minutes

at the factory before shipment. If you press the [Enter] or [Reject] key within the hold (output

stabilization) time, manual calibration is then completed.

8th Edition : Jan.13,2012-00