<10. Other Functions>

10-18

IM 11M12A01-02E

<Flow of span gas (air)>

The standard gas unit is used only when manual calibration is employed. Therefore, the timing

for flowing span gas (air) is included in the manual calibration flowchart described in Section

10.5.2. For operation of the converter, see Section 7.12, earlier in this manual.
(1) When the message “Open span gas valve” is displayed on the converter display during

calibration, plug the power cord into the power supply socket to start the pump of the

standard gas unit.

Enter

Open span gas valve.
Set flow span gas to
600ml/min.

Valve opened

Cancel calibration

Manual calibration

F10.15E.ai

Figure 10.15 Manual Calibration Display

(2) Next, adjust the flow rate to 600 ± 60 ml/min using the span gas valve “AIR” (the flow check

ball stops floating on the green line when the valve is slowly opened). To rotate the valve

shaft, loosen the lock nut and turn it using a flat-blade screwdriver. Turning the valve shaft

counterclockwise increases the flow rate.

(3) After adjusting the flow rate, tighten the valve lock nut.
(4) Select “Valve opened” (to start calibration) from the Manual calibration display shown in

Figure 10.15. Check the Trend graph display to see that the measured value is stabilized.

Then press the [Enter] key. The Manual calibration display shown in Figure 10.16 appears.

Disconnect the power cord to stop the pump.

Enter

Span calibration
Close the span gas valve.

Zero calibration

End

Manual calibration

F10.16E.ai

Figure 10.16 Manual Calibration Display

<Flow of zero gas>

Cause a zero gas to flow according to the Manual calibration display shown in Figure 10.17.

Enter

Open zero gas valve.
Set flow zero gas to
600ml/min.

Valve opened

Cancel calibration

Manual calibration

F10.17E.ai

Figure 10.17 Manual Calibration Display

8th Edition : Jan.13,2012-00

Advertising