<5. Wiring>

5-3

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

5.1.2

Wiring

Connect the following wiring to the converter. It requires a maximum of seven wiring connections

as shown below.
(1) Detector output (connects the converter with the detector.)
(2) Detector heater power (connects the converter with the detector.)
(3) Analog output signal
(4) Power and ground
(5) Contact output
(6) Operation of the solenoid valve of automatic calibration unit
(7) Contact input

1

FG

2

AO-1

(+)

3

AO-1

(-)

4

AO-2

(+)

5

AO-2

(-)

6

CJ

(+)

7

CJ

(-)

8

TC

(+)

9

TC

(-)

10

CELL

(+)

11

CELL

(-)

12

FG

13

DI-1

14

DI-2

15

DI-C

16

AI

(+)

17

AI

(-)

18

AC-

ZERO

19

AC-

SPAN

20

AC-

COM

21

FG

22

FG

CELL

(+)

CELL

(-)

TC

(+)

TC

(-)

CJ

(+)

CJ

(-)

23

DO-1

24

DO-1

25

DO-2

26

DO-2

27

DO-3

28

DO-3

29

DO-4

30

DO-4

31

HTR

32

HTR

33

L

34

N

35

G

36

FG

HTR HTR

100-240V AC

50/60 Hz

F5-3E.ai

ZR402G Separate type

Zirconia Oxygen Analyzer, Converter

ZR22G Separate type

Zirconia Oxygen Analyzer, Detector

Analog

output 2

Analog

output 1

4-20 mA DC

4-20 mA DC

Digital output

Contact

input 2

Contact input 1

Contact

output 1

Contact

output 2

Contact

output 3

Contact

output 4

Solenoid valve

for automatic calibration

Solenoid valve

for automatic calibration

Automatic Calibration Unit

ZR40H

Zero gas

Span gas

*1

*1

The ground wiring of the converter should be connected to either the protective ground terminal in the equipment or

the ground terminal of the converter case.

Ground to earth, ground resistance: 100 Ω or less.

*1

*1

Figure 5.2 Wiring connection to the converter